Synthetic Images

Region Size

Choosing an appropriate interrogation region or “window” size in PIV involves trade-offs. All other things equal, a larger window increases the number of particles per correlation and can increase the likelihood of finding a valid correlation. Increasing the window size, however, also decreases the spatial resolution of the displacement of the particles in each interrogation region (IR). This means that if the particle displacements vary within a given IR, the output vector may be less representative of the particles near the center of the IR (where the output vector is returned/placed).

A demonstration of these tradeoffs is shown in Figure 1. Imagine a pair of overlaid images with particle locations in the first image shown in red, and particle images in the second image shown in gold. To make the displacement of each particle easier to track, a black line is drawn from each particle location in the first image to the particle location in the second image.

Figure 1. A set of particle locations on two overlaid images. Image one locations are represented by the red dots, and image two locations are represented by the yellow dots. The black lines indicate that the movement from image one to image two of each particle.

Now, let’s compare three interrogation region sizes as shown in Figure 2.

Figure 2. Demonstration of the effects of varying window sizes (gray squares) on the resultant vector output (green arrow).

First, note that the IR shown in part A of Figure 2 is too small, since it has less than five particles, it has a higher likelihood (than an IR with more than five particles,) of particles exiting the IR and becoming “lost” between the first and second images. These lost particles are likely to lead to unreliable vector outputs. The vector output of small IRs, however, can represent the overall displacement near the center of the IR well, as long as it contains an adequate number of particles which appear in both images.

The IR shown in Part B of Figure 2 appears ideal, as it balances the need to capture an adequate number of particles between the first and second image to ensure a valid vector output located at the center of the correlation plane.

The IR shown in Part C of Figure 2 is too large. Although the number and percentage of particles present in both the first and second images is high (i.e., larger than five), the size of the IR results in the single vector output representing a large area. Using one vector output to represent the aggregate displacements of all particles contained in the large IR may provide unreliable results.

If desired, explore this by reducing the window size (with all other variables constant) to increase your resolution. However, be cautious of the output values as the window size and the number of particles in each region decreases. As well, the displacement between images relative to the IR size will increase. Reducing the window size too far can result in an unreliable (i.e., “spurious”) vector, and the error will be too great for this vector to be useful.

Author: Jack Elliott

Date Published: June, 2022